Preparing a CytoFLEX for Nanoscale Flow Cytometry
George Brittain, Sergei Gulnik, and Yong Chen, Beckman Coulter Life Sciences, Miami, FL 33196Introduction
Built around semiconductor technology, with a number of innovations to enhance light capture, reduce noise, and prevent signal losses, the CytoFLEX is capable of detecting biological nanoparticles (NPs) as small as 80nm by light scatter, and has a linear fluorescence range that extends down into the single digits for fluorophores like FITC. However, in order to properly setup the CytoFLEX for NP analyses, a variety of considerations need to be taken into account.
In this poster, we will demonstrate how to properly setup and clean a CytoFLEX flow cytometer for NP analyses. First, we will explore the different threshold options and sensitivity ranges. Next, we will show how to setup violet side scatter (VSSC) triggering. And finally, we will discuss several important issues that affect proper sample analyses, including how to clean the instrument to reduce noise within the nanoparticle range.
Materials

Methods
- Upon startup, the instrument was primed, cleaned and flushed.
- The beads were mixed together and diluted with HPLC water.
- Extracellular Vesicles (EVs) were prepared from fresh human blood as follows:
- 2mL of K3-EDTA blood was first aliquotted into 12x75mm centrifuge tubes. Cells in larger volumes, further away from the max radius, do not pellet as well within a short time frame and larger EVs pellet with longer time frames, so if a larger volume is needed, increasing the number of tubes works better than a greater volume per tube.
- The blood was centrifuged for 5 min at 200xg to pellet the majority of cells and platelets.
- Roughly 1mL of platelet-poor plasma (PPP) was removed from the top, careful to minimize collection of the platelet-rich plasma near the WBC layer.
- The PPP was filtered through a 200nm syringe filter to remove residual large particles.
- Finally, the filtered PPP was further purified using Izon size-exclusion columns to remove particles smaller than 70nm.
- All samples were acquired on a CytoFLEX-S N-V-B-R, and the data were analyzed in CytExpert v2.3.
I. Light-Scatter Sensitivity

Forward Scatter (FSC) on the CytoFLEX is not smallangle scatter, as found on a typical flow cytometer. It is a digital signal-analysis method, called axial light loss detection, which is optimized for 500nm-50m particles. It depends directly on the particle volume and is mostly independent of the refractive index.
Side Scatter (SSC) is much more sensitive, but is attenuated for use with cells and other large particles. It is optimized for roughly 200nm- 20µm.
Violet SSC (VSSC) is a 3rd scatter-detection mode that is available to take full advantage of the CytoFLEX sensitivity. VSSC can fully resolve 80nm PS or 100nm Si beads.
II. Setting up VSSC-H Triggering
- Set the Configuration to VSSC
- If a VSSC configuration does not exist, start a new one.
- Pick a channel and assign it as VSSC with the 405/10 filter.
- Save the configuration and set it as Current.

- Physically Swap the Filters
- Make sure that there is a filter in each position preceding the channels of interest or the light path will be broken.
- Do not place an equivalent bandpass filter upstream of a channel of interest or this will also result in the loss of the desired signal.

- Set the Threshold to VSSC-H
- VSSC-Height is the most sensitive trigger on the CytoFLEX for lightscatter detection. VSSC triggering requires 1, 2, and 3 together.
- The specific threshold level will depend on the VSSC Gain, and will change proportionally to any adjustments to the gain.
- The instrument always triggers in Height. Area back-calculates a relative Height setting, so it is more precise to set the Height directly.

- Change the Event Rate Setting to High
- If the Event Rate Setting is in Default, change it to High.
- The Default setting broadens the pulse window in order to help with setting up laser alignments and delays if they are out of line. The broader window will also increase the optical noise sampled.
- The High setting reduces background and improves event processing. The instrument should always be set on High for sample acquisition.

III. Important Considerations for Effective Nanoscale-Flow-Cytometry Experimentation
- Identify the Optimal Threshold
- Use appropriate reference particles to identify the optimal threshold. Using buffer alone will not properly identify the detection boundaries.
- On the CytoFLEX, the appropriate threshold level is generally around 10-fold higher than the gain. Above, the Gain = 400, Threshold = 3K.

- Clean the Sample Probe
- Debris from samples and buffers builds up in the probe, mostly visible below the 100nm PS range. Carryover can even occur between reads.
- Clean the probe by running a panel of Bleach, FlowClean, and then 2x Water to flush, for 1-2 min each at the max rate. Repeat if needed.

- Use Clean Buffers
- Any particulate in the sample buffers will also be detected as events.
- Filtering dirty buffers is generally insufficient because the particulate is often smaller than the filter. Discard and use fresh clean buffers.

- Optimize the Sample Concentration
- Flow cytometers have an optimal rate, inversely proportional to size. If the sample concentration is too high, swarming will occur.
- The concentration needs to be titrated to find the optimal range.
- At the appropriate concentration range, different dilutions will result in a shift up or down on the Counts axis, without shifting the intensity.
- The Abort Rate is also a swarming indicator: <5% is best.

- The Abort Rate Can Help Optimize
- The Abort Rate is particularly helpful with optimizing the dilutions of complex samples, e.g., without a tight monomodal distribution.
- When under 10%, the population medians should be consistent between dilutions. <5% is best.

- Balance the Concentration and Rate
- The sample rate is an instrument-based dilution directly into the flow cell. Faster rates proportionally increase the effective concentration.
- Faster rates also broaden the core stream and can increase CVs.

Discussion
Ultimately, the CytoFLEX is one of the most sensitive flow cytometers on the market. However, with such great power comes great responsibility to properly prepare the instrument and samples for effective nanoscale-flow-cytometry experiments. Nanoscale flow cytometry introduces a number of additional variables that are largely extraordinary to conventional flow cytometry, and each requires careful attention.
The CytoFLEX is for Research Use Only. The Beckman Coulter product and service marks mentioned herein are trademarks or registered trademarks of Beckman Coulter, Inc. in the United States and other countries. All other trademarks are the property of their respective owners.
Helpful Links
-
Дополнительные материалы
-
Рекомендации по применению
- 17-Marker, 18-Color Human Blood Phenotyping Made Easy with Flow Cytometry
- 21 CFR Part 11 Data Integrity for On-line WFI Instruments
- 8011+ Reporting Standards Feature and Synopsis
- Achieving Compliant Batch Release – Sterile Parenteral Quality Control
- Air Particle Monitoring ISO 21501-4 Impact
- Революция в аналитическом ультрацентрифугировании: новая центрифуга Optima AUC
- Analyzing Mussel/Mollusk Propagation using the Multisizer 4e Coulter Counter
- Automated 3D Cell Culture and Screening by Imaging and Flow Cytometry
- Автоматизированный анализ посева и роста клеток
- Automated Cell Transfection and Reporter Gene Assay
- Automated Cord Blood Cell Viability and Concentration Measurements Using the Vi‑CELL XR
- Ускорение ваших геномных исследований с автоматизацией пробоподготовки на Biomek
- Ускорение ваших геномных исследований с автоматизацией пробоподготовки на Biomek
- Ускорение исследований с автоматизированными станциями Biomek для геномики
- Automated salt-assisted liquid-liquid extraction
- Automated Sample Preparation for the Monitoring of Pharmaceutical and Illicit Drugs by LC-MS/MS
- Automated Transfection Methods
- Automated XTT Assay for Cell Viability Analysis
- Automating a Linear Density Gradient for Purification of a Protein:Ligand Complex
- Automating Biopharma Quality Control to Reduce Costs and Improve Data Integrity
- Automating Bradford Assays
- Automating Cell Line Development
- Automating Cell Line Development
- Automating Cell-Based Processes
- Automating the Cell Line Development Workflow
- The new Avanti J-15 Centrifuge Improves Sample Protection and Maximizes Sample Recovery
- The New Avanti J-15 Centrifuge Time Saving Deceleration Profile Improves Workflow Efficiency
- Комплексное решение для очистки белков с центрифугой Avanti JXN
- Avoid the Pitfalls When Automating Cell Viability Counting for Biopharmaceutical Quality Control
- Beer, Evaluation of Final Product and Filtration Efficiency
- Biomek Automated Genomic Sample Prep Accelerates Research
- Biomek Automated NGS Solutions Accelerate Genomic Research
- Preparation and purification of carbon nanotubes using an ultracentrifuge and automatic dispensing apparatus, and analysis using an analytical centrifuge system
- Cell Counting Performance of Vi–Cell BLU Cell Viability Analyzer
- Cell Line Development – Data Handling
- Cell Line Development – Hit Picking
- Cell Line Development – Limiting Dilution
- Cell Line Development – Selection and Enrichment
- Changes to GMP Force Cleanroom Re-Classifications
- Characterizing Insulin as a Biopharmaceutical Using Analytical Ultracentrifugation
- Clean Cabinet Air Particle Evaluation
- Cleanroom Routine Environmental Monitoring – FDA Guidance on 21 CFR Part 11 Data Integrity
- Comparing Data Quality & Optical Resolution of the Next Generation Optima AUC to the Proven ProteomeLab on a Model Protein System
- Conducting the ISO 14644-3 Cleanroom Recovery Test with the MET ONE 3400
- Considerations of Cell Counting Analysis when using Different Types of Cells
- Consistent Cell Maintenance and Plating through Automation
- Control Standards and Method Recommendations for the LS 13 320 XR
- Counting Efficiency: MET ONE Air Particle Counters and Compliance to ISO-21501
- Critical Particle Size Distribution for Cement using Laser Diffraction
- CytoFLEX
- Detecting Moisture in Hydraulic Fluid, Oil and Fuels
- Determination of Size and Concentration of Particles in Oils
- Efficient kit-free nucleic acid isolation uses a combination of precipitation and centrifugation separation methods
- dsDNA Quantification with the Echo 525 Liquid Handler for Miniaturized Reaction Volumes, Reduced Sample Input, and Cost Savings
- Compensation Setup For High Content DURAClone Reagents
- Echo System-Enhanced SMART-Seq v2 for RNA Sequencing
- Efficient Factorial Optimization of Transfection Conditions
- Решения для производства и разработки вакцин
- Enumeration And Size Distribution Of Yeast Cells In The Brewing Industry
- European Pharmacopoeia EP 2.2.44 and Total Organic Carbon
- Evaluation of Instrument to Instrument Performance of the Vi-CELL BLU Cell Viability Analyzer
- Exosome-Depleted FBS Using Beckman Coulter Centrifugation: The cost-effective, Consistent choice
- Flexible ELISA automation with the Biomek i5 Workstation
- Full Automation of the SISCAPA® Workflow using a Biomek NXP Laboratory Automation Workstation
- Fully-Automated Cellular Analysis by Flow Cytometry
- Соответствие требованиям GMP
- g-Max: Added Capabilities to Beckman Coulter's versatile Ultracentrifuge Line
- Grading of nanocellulose using a centrifuge
- A method of grading nanoparticles using ultracentrifugation in order to determine the accurate particle diameter
- Grading of pigment ink and measurement of particle diameter using ultracentrifugation / dynamic light scattering
- HIAC Industrial – Our overview solution for fluid power testing for all applications
- HIAC PODS+ Online Mode & Filter Cart Mode
- HIAC PODS+ versus Parker ACM-20 Performance comparison
- A complete workflow for high-throughput isolation of DNA and RNA from FFPE samples using Formapure XL Total on the KingFisher™ Sample Purification System: an application for robust and scalable cancer research and biomarker discovery
- A Highly Consistent BCA Assay on Biomek i-Series
- A Highly Consistent Bradford Assay on Biomek i-Series
- A Highly Consistent Lowry Method on Biomek i-Series
- Highly Reproducible Automated Proteomics Sample Preparation on Biomek i-Series
- Cell Line Development – Hit Picking
- How to Use Violet Laser Side Scatter Detect Nanoparticle
- How Violet Side Scatter Enables Nanoparticle Detection
- HRLD Recommended Volume Setting
- Hydraulic Oil Measurement
- ICH Q2 – the Challenge of Measuring Total Organic Carbon in Modern Pharmaceutical Water Systems
- ICH Q2 – The Challenge of Measuring Total Organic Carbon in Modern Pharmaceutical Water Systems
- Importance of TOC measurement in WFI in light of European Pharmacopoeia change
- Temperature dependence of hydrodynamic radius of an intrinsically disordered protein measured in the Optima AUC analytical ultracentrifuge.
- Isolation of cell-free DNA (cfDNA) from plasma using Apostle MiniMax™ High Efficiency cfDNA Isolation kit— comparison of fully automated, semi-automated and manual workflow processing
- Issues with Testing Jet Fuels for Contamination
- Leveraging the Vi-CELL MetaFLEX for Monitoring Cell Metabolic Activity
- Линейность результатов изучения олигомеризации БСА методом AUC с использованием оптики поглощения и интерференции
- Long Life Lasers
- LS 13 320 XR: Sample Preparation - How to measure success
- Particle Size Analysis Simple, Effective and Precise
- Matching Cell Counts between Vi–CELL XR and Vi–CELL BLU
- MET ONE Sensor Verification
- Metal colloid purification and concentration using ultracentrifugation
- Separation and purification of metal nanorods using density gradient centrifugation
- Method for Determining Cell Type Parameter Adjustment to Match Legacy Vi CELL XR
- Minimal Sample to Sample Carry Over with the HIAC 8011+
- Minimizing process variability in the manufacturing of bottled drinking water
- Modern Trends in Non‐Viable Particle Monitoring during Aseptic Processing
- Multi-Wavelength Analytical Ultracentrifugation of Human Serum Albumin complexed with Porphyrin
- Particle diameter measurement of a nanoparticle composite - Using density gradient ultracentrifugation and dynamic light scattering
- What to do now that ACFTD is discontinued
- Optimizing the HIAC 8011+ Particle Counter for Analyzing Viscous Fluids
- Optimizing the Multisizer 4e Coutler Counter for use with Small Apertures
- Optimizing Workflow Efficiency of Cleanroom Routine Environmental Monitoring
- Particle Counting in Mining Applications
- Particle testing in cleanroom high-pressure gas lines to ISO 14644 made easy with the MET ONE 3400 gas calibrations
- Pharma Manufacturing Environmental Monitoring
- Pharma Manufacturing Paperless Monitoring
- Принцип проточного центрифугирования
- Комплексное решение для очистки белков с центрифугой Avanti JXN
- Background Subtraction
- Calibrating the QbD1200 TOC Analyzer
- Detection Limit
- Inorganic Carbon Removal
- JP SDBS Validation
- Method Overview
- Overload Recovery
- QbD1200 Preparing Reagent Solution
- USP System Suitability
- Quality Control Electronic Records for 21 CFR part 11 Compliance
- Using the Coulter Principle to Quantify Particles in an Electrolytic Solution for Copper Acid Plating
- RCC A standardized automated approach for exosome isolation and characterization
- RCC Avanti J 15 Series Maximizing Sample Protection and Sample Recovery
- RCC Avanti J 15 Series Time Savings and Workflow Efficiency
- Root Cause Investigations for Pharmaceutical Water Systems
- Scalable Plasmid Purification using CosMCPrep
- Specification Comparison of Vi–CELL XR and Vi–CELL BLU
- Specifying Non-Viable Particle Monitoring for Aseptic Processing
- A Standardized, Automated Approach For Exosome Isolation And Characterization Using Beckman Coulter Instrumentation
- Switching from Oil Testing to Water and back using the HIAC 8011+ and HIAC PODS+
- Advanced analysis of human T cell subsets on the CytoFLEX flow cytometer using a 13 color tube-based DURAClone dry reagent
- Using k-Factor to Compare Rotor Efficiency
- USP 787 Small Volume Testing
- Validation of On-line Total Organic Carbon Analysers for Release Testing Using ICH Q2
- Vaporized Hydrogen Peroxide Decontamination of Vi–CELL BLU Instrument
- Vi CELL BLU FAST Mode Option
- Vi-CELL BLU Regulatory Compliance - 21 CFR Part 11
- A workflow for medium-throughput isolation of cfDNA from plasma samples using Apostle MiniMax™ on the KingFisher™ Technology
- Брошюры, буклеты и проспекты
- Каталоги
- Буклеты
-
Постеры
- Применение ультрацентрифугирования для очистки и характеризации биомолекул
- Automating Genomic DNA Extraction from Whole Blood and Serum with GenFind V3 on the Biomek i7 Hybrid Genomic Workstation
- ABRF 2019: Automated Genomic DNA Extraction from Large Volume Whole Blood
- AACR 2019: Isolation and Separation of DNA and RNA from a Single Tissue or Cell Culture Sample
- AACR 2019: Correlation between Mutations Found in FFPE Tumor Tissue and Paired cfDNA Samples
- AGBT 2019: A Scalable and Automatable Method for the Extraction of cfDNA
- ABRF 2019: Simultaneous DNA and RNA Extraction from Formalin-Fixed Paraffin Embedded (FFPE) Tissue
- AMP 2019: Correlation Between Mutations Found in FFPE Tumor Tissue and Paired cfDNA Samples
- Automated library preparation for the MCI Advantage Cancer Panel at Miami Cancer Institute utilizing the Beckman Coulter Biomek i5 Span-8 NGS Workstation
- Automating Cell Line Development for Biologics
- Cellular Challenges: Taking an Aim at Cancer
- Cell-Line Engineering
- Characterizing the Light-Scatter Sensitivity of the CytoFLEX Flow Cytometer
- ASHG 2019: Comparison between Mutation Profiles of Paired Whole Blood and cfDNA Samples
- ASHG 2019: Correlation Between Mutations Found in FFPE Tumor Tissue and Paired cfDNA Samples
- Mastering Cell Counting
- Preparing a CytoFLEX for Nanoscale Flow Cytometry
- A Prototype CytoFLEX for High-Sensitivity, Multiparametric Nanoparticle Analysis
- Пошаговые руководства
- Протоколы
-
Практические примеры
- Adenoviral Vectors Preparation
- Algae Biofuel Production
- Autophagy
- B Cell Research
- Cardiovascular Disease Research
- Cell Marker Analysis
- Choosing a Tabletop Centrifuge
- Collagen Disease Treatment
- Controlling Immune Response
- Creating Therapeutic Agents
- DNA Extraction from FFPE Tissue
- English Safety Seminar
- Equipment Management
- Exosome Purification Separation
- Future of Fishing Immune Research
- Hematopoietic Tumor Cells
- iPS Cell Research
- Membrane Protein Purification X Ray Crystallography
- Organelles Simple Fractionation
- Particle Interaction
- Retinal Cell Regeneration
- Sedimentary Geology
- Severe Liver Disease Treatment
- Treating Cirrhosis
- University Equipment Management
- Fundamentals of Ultracentrifugal Virus Purification
-
Публикации
- Centrifugation is a complete workflow solution for protein purification and protein aggregation quantification
- AUC Insights - Analysis of Protein-Protein-Interactions by Analytical Ultracentrifugation
- AUC Insights - Assessing the quality of adeno-associated virus gene therapy vectors by sedimentation velocity analysis
- AUC Insights - Sample concentration in the Analytical Ultracentrifuge AUC and the relevance of AUC data for the mass of complexes, aggregation content and association constants
- Hydraulic Particle Counter Sample Preparation
- Электронные книги
- Interviews
-
Рекомендации по применению