Avanti® J-30I Centrifuge, 3-Phase, 50 Hz, 380 V

363121


Enhanced performance compared to similar high speed or super speed centrifuges. New rotors enable forces of over 110,000g. Features include a high-torque SR Drive for the fastest acceleration and deceleration possible with gentle precision, automatic rotor identification and imbalance safety check, CFC-free refrigerants, and a friction reduction system. Compatible with all Beckman Coulter J rotors. System offers the fastest separations in the shortest time. Swinging-bucket and fixed-angle rotors provide maximum separation forces beyond 100,000 x g at speeds up to 30,000 rpm. Your productivity is ensured by unmatched accel/decel rates and an easy run set-up. Your sample spends more time at full force and less time in the centrifuge. Operates at near-Ultracentrifuge forces at over 100,000 x g for fast protein separations in the Avanti J-30I and the JA-30.50, the highest g-force fixed-angle rotor available in a high-performance centrifuge. Subcellular organelles are effectively separated by rate zonal centrifugation with the JS-24.38, a 6 x 38 mL swinging-bucket rotor. High Throughput: Four-liter batch throughput for bacteria and cell membrane isolation using the JLA-9.1000 J-LITE rotor at 16,800 x g. DNA sample preparation in up to 10 microplates with the JS-5.9 Rotor. Exclusive SR Drive technology offers the fastest accel and decel times. Low heat output and low energy consumption in a high efficiency machine.

Технические спецификации

Platform Avanti
Maximum Speed 30,000 rpm
Maximum g-Force 110,500.00|x g||
BioSafe Нет
Acceleration Maximum, Slow or Timed (1-10 minutes, 0-500 rpm)
Deceleration Maximum, Slow, Timed (1-10 minutes from 500-0 rpm) or Off
Speed Control ± 10 rpm of set speed or 0.15%, whichever is greater
Timed Run 1 minute-99 hours, 59 minutes; Hold or Ω 2t
Max Capacity Volume Range 1 - 4 L
Refrigeration System Non-CFC, non-ozone depleting refrigerant
Drive Type Cooling SR Drive* / Air cooled
Temperature Range(s) -20° to +40° C
Rotor ID Automatic
Friction Reduction System Yes
Zonal Capability Да
Pollution Degree 2
Imbalance Tolerance ±
Environmental Conditions -20°C to 40°C in 1° increments
Installation Category II
Software Virtual Network Computing (VNC), Mobile Fuge app for Apple iOS and Android
Material(s) Steel, Structural Foam
Finish Powder Coat Paint
Weight 310 kg / 683.4 lb
Width 71 cm / 28 in
Depth 86 cm (102.0 cm with air direviter extending)
Height 86 cm (141 cm with door open)
Clearances 7.6 cm (sides) & 16 cm (rear)
Humidity Restrictions < 95% (non-condensing)
Power Requirements 220/380 VAC, 50 Hz, 3 Phase

Посетители также интересовались

НЕ ВСЕ ПРОДУКТЫ ДОСТУПНЫ ВО ВСЕХ СТРАНАХ
ДОСТУПНОСТЬ И РЕГУЛЯТОРНЫЙ СТАТУС ПРОДУКЦИИ ЗАВИСИТ ОТ РЕГИСТРАЦИИ ПРОДУКТА В СООТВЕТСТВИИ С МЕСТНЫМ ЗАКОНОДАТЕЛЬСТВОМ

Что означает регулярный статус продукта:
IVD: In Vitro Diagnostic. Данная продукция предназначена для клинической диагностики in vitro.
ASR: Analyte Specific Reagents. Специфические аналитические реагенты. Аналитические и технические характеристики не установлены.
CE: Продукция, предназначенная для клинической диагностики in vitro и соответствующая директиве Европейского союза (98/79/EC). (Примечание: Оборудование может соответствовать другим директивам Европейского союза, и иметь соответствующую маркировку.)
RUO: Research Use Only. Данная продукция предназначена только для научных исследований и не предназначена для клинической диагностики.
LUO: Laboratory Use Only. Продукция предназначены только для лабораторного использования.
Без регуляторного статуса: немедицинские изделия и изделия, не подлежащие обязательной регистрации в регуляторных органах. Данная продукция не предназначена для клинического применения.